How Work Flame Ionization Detection (FID) in TVA-1000B?

Flame Ionization Detection (FID) A Flame Ionization Detector (FID) measures organic compounds by utilizing a flame produced  by the combustion of hydrogen and air. When hydrocarbons in the sample are introduced  to the detection zone, ions are produced by the following reaction: RH + O → RHO+ + e– → H2O + CO2 where R = carbon compound A collector electrode with a polarizing voltage is also located within the detector chamber, and  the ions produced by this reaction are attracted to it. As the ions migrate towards the collector, a current is produced which is directly proportional to the concentration of hydrocarbons  introduced to the flame. This current is then amplified and sent to a microprocessor and/or  analog readout device. The FID has a wide dynamic range. The effective dynamic range can be further expanded by  use of a dilutor kit which reduces very high volatile organic compounds (VOC) concentrations  to within the dynamic range (or even line...